Page 1 ABX Micros ES60 User manual P/n: RAB237AEN Explore the future.; Page 3 ABX MICROS ES60 OT/CT User Manual P/n: RAB237AEN HORIBA ABX SAS B.P. 7290 34184 MONTPELLIER Cedex 4 - FRANCE.; Page 4 When a subsequent software version is released, only the electronic version (CD-ROM and/or online help) of this user manual is updated and supplied by HORIBA ABX. Horiba Abx Micros 60 in Lab Equipment And Accessories for sale used HORIBA ABX Micros 60 for HORIBA ABX Micros HORIBA ABX Micros HORIBA ABX Micros 60: pin. Abx micros 60 manual ABX Micros 60: pin. CDS/ABX Micros 45/60 Micros Diluent CDS/ABX Micros 45/60 Micros Diluent CDS-501-128 (10 Liters) CDS/Micros Diluent (10 Liters) for use with. More ABX Micros 60/ABX Micros ES 60 Reagents More CellaVision® Digital Cell Morphology Systems. More Hematovision Online Cytology Software. More ISO 15189 Accreditation Support. More QCP Quality Control Program. More QSP2.0 Quality Slide Program. Related products. BY SAME TECHNIQUE.
- Documents
- Videos
- News
ABX Micros ES60 Hematology Analyzer, a new generation hematology analyzer based on the Micros concept. It is universally recognized for its reliability, robustness and high quality results. Features: - 60 samples per hour (OT Version) - CBC + 3 DIFF (18 parameters) - Large color touch screen. The ABX MICROS 60 should be placed on a clean and level table or workbench. Please note that the ABX MICROS 60, printer, and reagent weights are approximately 30 kilograms (66 lbs.). Avoid exposure to sunlight. Proper ventilation requires adequate space behind the instrument. At least 20 cm (8 inches) must be maintained behind the instrument.
DESCRIPTION
8 or 18 fully automated parametersCBC / 3 Part DIFFOpen or Closed TubeEasy pose for mac os. Impedance technology, photometry
Microsampling: Only 10 µL
Smart card based memory storage (optional)
No compressor (stepper motor technology)
FORUMSView All (20)
Ask a New Question2Replies5 months ago | 5 months agoHavingproblem with RDW, Low results Having problems with low RDW results.RBC 5,09-HGB 15,1-HCT 49,3 MCV 97-MCH 29,7 MCHC 30,66 PLT 230 RDW 2,2. Reply |
-Mkama zebedayo a year ago | a year agoservicing micro 60 and screen master i would like to know what can i do to my micro 60 that is giving out high PLT reading that is going up to 22222 sometimes giving DIL.also to my screen master that is giving out a wide range of deviation to a single sample or QC testing sample aslo the failure of QC,How to calibrate this one since it has failed more than fifth times?Reply |
2Replies2 years ago | 2 years agoSERVICE MANUAL Dear All,Am looking for a service manual for Horiba ABX Micros 60. May anyone who has it share by sending to: vgmakwinja@medcol.mwThanks in advance.Reply |
SERVICE COMPANIESView All Hematology Analyzer Companies
FEATURES
Measurement methods: Impedance technology, photometry. Liquid valves (without pinching): Reliability and increased precision No compressor (stepper motor): Reliability, less maintenance and noise
Additional Specifications
Technical
Weight
31 lbs
Power supply
90/240 V - 50/60 Hz - 230 VA.
Dimensions
16.5 in x 14 x 12.5 in (HxWxD).
Printer
Dot matrix.
Throughout
60 samples/hour (on open tube)
55 samples/hour (closed tube)
Data management
Automated smart cards used for QC and patient data storage (optional).
RS 232 output.
Mono-directional connection.
Memory
Last result (with curves) or 78 results (optional smart cards based memory)
Screen
2 lines of 40 alphanumerical characters.
Keyboard
Barcode reader (optional).
Samples
60 samples/hour (open tube)
Parameters (16+2)
CBC Mode (18 parameters): WBC, RBC, HGB, HCT, MCV,*RDW, MCH, MCHC, PLT, MPV, PCT, *PDW, % and # of LYM, MON and GRA
Graphics
LMG histogram, RBC, PLT histograms
Sample-Volume
CBC mode : 10 µL
Stability
24 hours (CBC)
Operating sample mode
Open tube mode with 4 position tube holder (QC, 125 or 500 µL, 3 or 5 mL)
Identification
Patient I.D., Sequence#
Flags
Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags
* RUO (Research Use Only)
Page 1 ABX Micros ES60 User manual P/n: RAB237AEN Explore the future.; Page 3 ABX MICROS ES60 OT/CT User Manual P/n: RAB237AEN HORIBA ABX SAS B.P. 7290 34184 MONTPELLIER Cedex 4 - FRANCE.; Page 4 When a subsequent software version is released, only the electronic version (CD-ROM and/or online help) of this user manual is updated and supplied by HORIBA ABX. Horiba Abx Micros 60 in Lab Equipment And Accessories for sale used HORIBA ABX Micros 60 for HORIBA ABX Micros HORIBA ABX Micros HORIBA ABX Micros 60: pin. Abx micros 60 manual ABX Micros 60: pin. CDS/ABX Micros 45/60 Micros Diluent CDS/ABX Micros 45/60 Micros Diluent CDS-501-128 (10 Liters) CDS/Micros Diluent (10 Liters) for use with. More ABX Micros 60/ABX Micros ES 60 Reagents More CellaVision® Digital Cell Morphology Systems. More Hematovision Online Cytology Software. More ISO 15189 Accreditation Support. More QCP Quality Control Program. More QSP2.0 Quality Slide Program. Related products. BY SAME TECHNIQUE.
- Documents
- Videos
- News
ABX Micros ES60 Hematology Analyzer, a new generation hematology analyzer based on the Micros concept. It is universally recognized for its reliability, robustness and high quality results. Features: - 60 samples per hour (OT Version) - CBC + 3 DIFF (18 parameters) - Large color touch screen. The ABX MICROS 60 should be placed on a clean and level table or workbench. Please note that the ABX MICROS 60, printer, and reagent weights are approximately 30 kilograms (66 lbs.). Avoid exposure to sunlight. Proper ventilation requires adequate space behind the instrument. At least 20 cm (8 inches) must be maintained behind the instrument.
DESCRIPTION
8 or 18 fully automated parametersCBC / 3 Part DIFFOpen or Closed TubeEasy pose for mac os. Impedance technology, photometry
Microsampling: Only 10 µL
Smart card based memory storage (optional)
No compressor (stepper motor technology)
FORUMSView All (20)
Ask a New Question2Replies5 months ago | 5 months agoHavingproblem with RDW, Low results Having problems with low RDW results.RBC 5,09-HGB 15,1-HCT 49,3 MCV 97-MCH 29,7 MCHC 30,66 PLT 230 RDW 2,2. Reply |
-Mkama zebedayo a year ago | a year agoservicing micro 60 and screen master i would like to know what can i do to my micro 60 that is giving out high PLT reading that is going up to 22222 sometimes giving DIL.also to my screen master that is giving out a wide range of deviation to a single sample or QC testing sample aslo the failure of QC,How to calibrate this one since it has failed more than fifth times?Reply |
2Replies2 years ago | 2 years agoSERVICE MANUAL Dear All,Am looking for a service manual for Horiba ABX Micros 60. May anyone who has it share by sending to: vgmakwinja@medcol.mwThanks in advance.Reply |
SERVICE COMPANIESView All Hematology Analyzer Companies
FEATURES
Measurement methods: Impedance technology, photometry. Liquid valves (without pinching): Reliability and increased precision No compressor (stepper motor): Reliability, less maintenance and noise
Additional Specifications
Technical
Weight
31 lbs
Power supply
90/240 V - 50/60 Hz - 230 VA.
Dimensions
16.5 in x 14 x 12.5 in (HxWxD).
Printer
Dot matrix.
Throughout
60 samples/hour (on open tube)
55 samples/hour (closed tube)
Data management
Automated smart cards used for QC and patient data storage (optional).
RS 232 output.
Mono-directional connection.
Memory
Last result (with curves) or 78 results (optional smart cards based memory)
Screen
2 lines of 40 alphanumerical characters.
Keyboard
Barcode reader (optional).
Samples
60 samples/hour (open tube)
Parameters (16+2)
CBC Mode (18 parameters): WBC, RBC, HGB, HCT, MCV,*RDW, MCH, MCHC, PLT, MPV, PCT, *PDW, % and # of LYM, MON and GRA
Graphics
LMG histogram, RBC, PLT histograms
Sample-Volume
CBC mode : 10 µL
Stability
24 hours (CBC)
Operating sample mode
Open tube mode with 4 position tube holder (QC, 125 or 500 µL, 3 or 5 mL)
Identification
Patient I.D., Sequence#
Flags
Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags
* RUO (Research Use Only)
- Documents
- Videos
- News
DESCRIPTION
8 or 18 fully automated parametersCBC / 3 Part DIFFOpen or Closed TubeImpedance technology, photometry
Microsampling: Only 10 µL
Smart card based memory storage (optional)
No compressor (stepper motor technology)
FORUMSView All (20)
Ask a New QuestionHoriba Micros 60 User Manual
2Replies5 months ago | 5 months agoHavingproblem with RDW, Low results Having problems with low RDW results.RBC 5,09-HGB 15,1-HCT 49,3 MCV 97-MCH 29,7 MCHC 30,66 PLT 230 RDW 2,2. Reply |
-Mkama zebedayo a year ago | a year agoservicing micro 60 and screen master i would like to know what can i do to my micro 60 that is giving out high PLT reading that is going up to 22222 sometimes giving DIL.also to my screen master that is giving out a wide range of deviation to a single sample or QC testing sample aslo the failure of QC,How to calibrate this one since it has failed more than fifth times?Reply |
2Replies2 years ago | 2 years agoSERVICE MANUAL Dear All,Am looking for a service manual for Horiba ABX Micros 60. May anyone who has it share by sending to: vgmakwinja@medcol.mwThanks in advance.Reply |
SERVICE COMPANIESView All Hematology Analyzer Companies
FEATURES
Abx Pentra 60
Measurement methods: Impedance technology, photometry. Liquid valves (without pinching): Reliability and increased precision No compressor (stepper motor): Reliability, less maintenance and noise
Abx Micros Es 60 User Manual
Additional Specifications
Abx Micros 60 - Manual Pdf
Technical
Weight
31 lbs
Power supply
90/240 V - 50/60 Hz - 230 VA.
Dimensions
16.5 in x 14 x 12.5 in (HxWxD).
Printer
Dot matrix.
Throughout
60 samples/hour (on open tube)
55 samples/hour (closed tube)
Data management
Automated smart cards used for QC and patient data storage (optional).
RS 232 output.
Mono-directional connection.
Memory
Last result (with curves) or 78 results (optional smart cards based memory)
Screen
2 lines of 40 alphanumerical characters.
Keyboard
Barcode reader (optional).
Samples
60 samples/hour (open tube)
Parameters (16+2)
CBC Mode (18 parameters): WBC, RBC, HGB, HCT, MCV,*RDW, MCH, MCHC, PLT, MPV, PCT, *PDW, % and # of LYM, MON and GRA
Graphics
LMG histogram, RBC, PLT histograms
Sample-Volume
CBC mode : 10 µL
Stability
24 hours (CBC)
Operating sample mode
Open tube mode with 4 position tube holder (QC, 125 or 500 µL, 3 or 5 mL)
Identification
Patient I.D., Sequence#
Flags
Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags
* RUO (Research Use Only)